Analyte's separated by HILIC vs Reverse phase Chromatography (RPC), using increasingly organic and aqueous mobile phases respectively. Analytes organized by lipophilicity.
Hydrophilic interaction chromatography (or hydrophilic interaction liquid chromatography, HILIC) is a type of liquid chromatography that uses a hydrophilic stationary phase and a high-organic mobile phase for the separation of analytes by polarity.[1][2] While it is not as popular as some other types of liquid chromatography, the number of scientific publications using HILIC have greatly increased since the early 2000s.[3] HILIC is similar to reverse phase chromatography in its mobile phase composition, and also to normal phase chromatography, with its polar stationary phase.[4][5] It also has overlap with ion exchange chromatography.[4] Sometimes, HILIC is considered to be a hybrid of these techniques.[6]
HILIC was named in 1990 by Andrew Alpert, who described it as a type of liquid-liquid partition chromatography.[2] He suggested that analytes elute in order of increasing polarity,[2] a conclusion supported by review and re-evaluation of published data.[7] The mechanism for HILIC is still not entirely understood, but it is thought to rely on analytes partitioning between the organic-rich mobile phase and a water-enriched layer that forms of the surface of the polar stationary phase, in a liquid-liquid extraction system.[2][5] More polar analytes will have stronger interactions with the water-enriched layer and with the column itself, therefore being retained on the column for longer.[3]
One of the key factors influencing HILIC separations is the chemical nature of the stationary phase that is packed into the column.[8] Stationary phases on HILIC columns not only provide physical support for the water layer which analytes separate into, but also interact with the analytes through hydrogen bonding and electrostatic interactions, affecting their retention and therefore the mechanism of separation.[3][5]
Typical HILIC stationary phases are polar, made of classical bare silica or silica gels modified with various polar groups.[1][9] Some commonly used stationary phases include bare silica, or silica chemically bonded to amino-,[10] amide-,[11] cyano-, or diol- groups.[12][9] Ion exchanger groups, both cationic and anionic {{citation needed|date=November 2025} 6">Lardeux, Honorine; Guillarme, Davy; D'Atri, Valentina (2023-02-08). "Comprehensive evaluation of zwitterionic hydrophilic liquid chromatography stationary phases for oligonucleotide characterization". Journal of Chromatography A1690. doi:10.1016/j.chroma.2023.463785. ISSN0021-9673. PMID36641941.</ref> groups are also commonly used.[9]
While most HILIC phases are polar, there have also been exceptions where non-polar bonded silicas are used with extremely high organic solvent composition. In this case, interactions are affected by exposed patches of silica in between the bonded ligands on the support.[13]
Mobile phase
The mobile phase, or the liquid phase that runs across the column during separation, for HILIC is typically composed of a high amount of water-miscible, polar organic solvent and a low amount of water.[6] Typically, acetonitrile ("MeCN", also designated as "ACN") is used for the organic solvent, though other aprotic water-miscible solvents, such as alcohols at higher concentration, tetrahydrofuran, or dioxane, can also be used.[3]
As with other methods of chromatography, the mobile phase can be delivered isocratically or with a gradient starting at high-organic progressing towards increasing aqueous content.[3] If using a mobile phase gradient, the mobile phase will progressively increases in polar-aqueous content, causing increasingly polar analytes to be eluted.[2][14]
Additives
Mobile phase pH and electrostatic interactions, as well as analyte polarity, are regulated by the addition of ionic additives.[3] These additives improve separation efficiency, including more symmetric peaks, less peak tailing, and better recovery from the stationary phase.[3][15]Ammonium acetate and ammonium formate are commonly used, as they have good solubility in high organic.[16]
When considering additive addition, compatibility with detectors is important to consider. HILIC is often used with a mass spectrometry (MS), which cannot handle non-volatile salts like sodium perchlorate, which may suppress ion signal in the instrument, though it may increase mobile phase polarity and assist with elution in HILIC.[3][17]
Choice of pH
With surface chemistries that are weakly ionic, the choice of pH can affect the ionic nature of the column chemistry. Properly adjusted, the pH can be set to reduce the selectivity toward functional groups with the same charge as the column, or enhance it for oppositely charged functional groups.{{citation needed|date=November 2025} the polarity of the solutes. However, for column surface chemistries that are strongly ionic, and thus resistant to pH values in the mid-range of the pH scale (pH 3.5–8.5), these separations will be reflective of the polarity of the analytes alone.[citation needed] When used with mass spectrometry, p nd thus detection.[18]
ERLIC
ERLIC (electrostatic repulsion interaction chromatography) is a type of HILIC that relies on electrostatic interactions, coined by Alpert in 2008.[19] The ionic stationary phase in ERLIC is chosen to have a similar charge to the analyte(s) so that the analyte is repelled by the stationary phase but also retained by the aqueous layer, allowing for enhanced interaction of the remaining polar, oppositely-charged functional groups of the analyte.[20][19] Electrostatic effects have an order of magnitude stronger chemical potential than neutral polar effects. These opposing effects can, in some cases, enable isocratic separations, with the mobile phase held constant instead of delivered at a gradient.[20] ERLIC can be used to reduce retention of more polar functional groups and minimize the influence of common ionic groups within a set of analytes. [citation needed]
Similarly, a positively charged anion exchange column can be used to reduce the influence of cationic (positively charged) functional groups on the retention time of analytes. For example, when selectively isolating phosphorylated peptides or sulfated polysaccharide molecules, use of a pH between 1 and 2 pH units reduces the polarity of two of the three ionizable oxygens of the phosphate group, and thus allows easy desorption from the (oppositely charged) surface chemistry. Negatively charged carboxyl groups in the analyte will be protonated at this low pH, and thus also contribute less to the polarity and therefore separation of the analyte {{citation needed|date=November 2025} umb|332x332px|Comparing HILIC, normal phase, and reversed phase chromatography in terms of sensitivity for electrospray ionization mass spectrometry and analyte polarity. ]]
Applications
HILIC can be applied in many fields including proteomics,[20]metabolomics,[21] medical studies,[22][23] and agricultural/ food studies,[24] among others. It can be used to separate proteins and peptides, nucleosides, amino acids, sacharides, carbohydrates, and other small, polar, ionizable compounds.[25] HILIC is especially common in metabolomic studies, both for targeted and untargeted approaches, given its ability to retain polar analytes that are poorly suited for traditional reverse-phased columns.[21][2] This separation technique is also particularly suitable for glycosylation analysis[26] and quality assurance of glycoproteins and glycoforms in biologic medical products.[27] For the detection of polar compounds with the use of electrospray-ionization mass spectrometry as a chromatographic detector, HILIC can offer a ten fold increase in sensitivity over reversed-phase chromatography because the organic solvent is much more volatile.[28] HILIC is considered orthogonal reverse phase chromatography, and these two are increasingly being combined in studies for more comprehensive coverage.[29][30]
↑ 5.05.15.2Redón, Lídia; Subirats, Xavier; Rosés, Martí (2021-10-25). "Volume and composition of semi-adsorbed stationary phases in hydrophilic interaction liquid chromatography. Comparison of water adsorption in common stationary phases and eluents". Journal of Chromatography A1656. doi:10.1016/j.chroma.2021.462543. ISSN0021-9673. PMID34571282.