}} Ghosting is a visual artifact that occurs in magnetic resonance imaging (MRI) scans. This artifact can be a consequence of environmental factors or the human body (such as blood flow, implants etc.). Ghosting is a multidimensional artifact that occurs in the MRI in the phase encoded direction (short axis of the image) after applying the Fourier transform.
When the phase of the magnetic resonance signal is being encoded into the 2-D or 3-D Fourier image, a mild deviation from the actual phase and amplitude may occur. This incompatibility of parameters causes ghosting. The reasons this occurs are often physical factors, such as temperature or humidity of the environment, or movement caused by the patient or in the patient’s body. There are two types of movement that could result in ghosting:
These movements, when reflected over the anatomy may result in improper diagnosis of the disease and hence have to be identified and suppressed.
In MRI, phase encoding is the process of acquiring data by altering the phase of the spin of the atom by applying magnetic pulses before acquiring the actual data.
k-space is a graphic matrix of an MR that represents the Fourier transform domain of an image before undergoing the Inverse Fourier transformation.The phase deviations that occur in the k-space of an MR image decide the characteristics of the ghosts that appear in the resulting image. Even though most of the ghosting is due to the phase deviation in the phase encoded direction, they also appear in other directions of the k-space.
The basic principle of ghosting can be explained with the help of Echo-Planar Imaging (EPI). Echo-planar imaging is an MRI technique that reduces the time of data acquisition to reduce capture of patient movement. An image in the EPI can be captured in between 20-100 milliseconds. Multiple lines of data are created by transmitting RF pulse sequences with a gradient difference of 90° and 180°. After the 180° pulse, the frequency encoding gradient rapidly changes to a negative amplitude and the resulting echos are encoded in the phase encoded axis. The pulses that are used to excite can be classified into 2 types, namely 'single shot' and 'multi-shot' pulse sequences. The multi-shot echo planar images tend to capture more data than that of the single shot EPI.
The process of removing the movement of images in an MRI scan is known as de-ghosting.
Several algorithms have been proposed to remove ghosting in the medical images.
The iterative problem solving method[1] is a ghost correction algorithm that removes ghosting that occurs due to the physical movement of the patient. This is a post-processing technique which uses the simple motion models (such as translational motion, rotational motion or linear motion) to remove the ghosts that occur in the MR images. This algorithm uses an iterative approach to correct the distorted image by using the motion models.
In a standard rectangular-grid acquisition system, each row of data is acquired by applying a gradient [math]\displaystyle{ G_y }[/math] in the direction of the y-axis with a fixed time T before the data is acquired. The process of acquiring data in a row is known as view. This data is scanned by encoding in the frequency encoding and phase encoding directions. This is followed by taking the Inverse Fourier transform for reconstruction of the image, which can contain ghosting artifacts. The iterative method is then applied to reduce the ghosting artifacts.
As this is a post-processing technique and requires the pre-defined model of the motion, the inter-view motion detected signal is compared with the existing models. The theoretically generated magnetization of the image is calculated. This magnetization should match with the magnetization that is observed.
Iterative inverse problem solving is faster than Cardiac gating[2] and doesn't involve the patient too much.
The choice of the motion model is critical, as it should be sufficiently close to the actual model.
Reference free EPI ghost correction algorithm[3] uses a method called ALOHA (Annihilating filter-based low rank Hankel structured matrix completion approach). The data of the k-space matrix is numbered consecutively and is split into odd and even data based on the samples of the index. This method was developed based on the fact that the difference between the odd and even virtual k-space data is the Fourier transform of the underlying sparse image. It is based on the principle that Ghost Nyquists are produced due to the inconsistencies between the odd and even echos of the MR images. The occurrence of the Ghosts in the images is converted into missing k-space data and is recovered with the help of the ALOHA matrix.
There are 2 types of approaches that are used to remove the Nyquist Ghost artifacts:
The odd and the even samples of data are taken from the k-space by means of interpolation. A high-performance interpolation method would be able to find the missing data from the actual even and odd sequences and remove the ghosting artifacts from the images. ALOHA converts the sparse recovery problem into a k-space interpolation problem using a low-rank interpolator.
Reference free algorithms are prone to errors and have low performance when compared to the reference based algorithms.
The method of generalized projections (MGP) is based on the fact that discontinuities in the phase form alternating patterns in the k-space, which leads to ghosting. Interleaved EPI gives a train of ghosts in the phase encoded direction. This EPI ghosting can be corrected from the phase using 2 or more reference scans.
According to MGP, the region of support (ROS) is defined around the parent image (the object in the image that is ghosted) of the ghosted image. The pixels outside the ROS is masked to zero. The Fourier transform of the masked image is taken. The phase of the original image is changed to the model image (the image that is modeled as ghosting arising from the phase offsets) but the magnitude is kept constant. The assumption in the MGP method is that the model k-space phase of each line is more precise than that of the original k-space image. From the MGP algorithm of ghost correction, it has been observed that it works best only when there is low degree of freedom. This method can be applied only when no more than 8 parameters are required to be found.
The projections on the slices have their own magnetization values. These different values of magnetization will induce a different phase response. Hence a new measurement of the phase response has to be taken.
[7] [8] [9] [10] [11] [12] [13]
Original source: https://en.wikipedia.org/wiki/Ghosting (medical imaging).